Effect of femoral head size on wear of the polyethylene acetabular component.

J Bone Joint Surg Am · 1990 · Classic · cited 1,042 times

J Livermore, D M Ilstrup, B.F. Morrey

Adult Reconstruction

SUMMARY — THE REDUCTIONIn 385 hips followed over 9.5 years, 28-mm femoral heads produced the least linear polyethylene wear, 32-mm heads the most volumetric wear, and greater wear correlated with proximal femoral resorption and lysis.
Abstract, as published

A technique was developed to determine the wear of the acetabular component of a total hip replacement by examination of standardized initial and follow-up radiographs. Three hundred and eighty-five hips were followed for at least 9.5 years after replacement. The least amount and rate of linear wear were associated with use of a femoral head that had a diameter of twenty-eight millimeters (p less than 0.001). The greatest amount and mean rate of linear wear occurred with twenty-two-millimeter components, but these differences were not statistically significant. The greatest volumetric wear and mean rate rate of volumetric wear were seen with thirty-two-millimeter components (p less than 0.001). A wider radiolucent line in acetabular Zone 1 was associated with use of the thirty-two-millimeter head. The amounts of resorption of the proximal part of the femoral neck and of lysis of the proximal part of the femur both correlated positively with the extent of linear and volumetric wear; this suggests an association between the amount of debris from wear and these changes in the femoral neck and proximal part of the femur.

Featured in the 2026-10-01 issue.

← Identification of patient subgroups by hierarchical cluster a…Are We Winning or Losing the Battle With Periprosthetic Joint… →

The Reduction is a free email digest of newly published orthopaedic literature — a handful of new papers in the subspecialties you choose, each summarized like this one. Subscribe free or browse the archive.